Title of article :
Calibration of CMD-2 drift chamber
Author/Authors :
Anashkin، نويسنده , , E.V. and Chernyak، نويسنده , , D.V and Fedotovich، نويسنده , , G.V and Khazin، نويسنده , , B.I and Lukin، نويسنده , , P.A and Maksimov، نويسنده , , A.V and Ruban، نويسنده , , A.A and Snopkov، نويسنده , , I.G and Solodov، نويسنده , , E.P and Zavarzin، نويسنده , , V.G، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1996
Pages :
2
From page :
432
To page :
433
Abstract :
The experience gained during four years of operation of the CMD-2 drift chamber is reported. Offline calibration of the DC parameters using a sample of Bhabha scattering events is described. Fine corrections to the Lorentz angle allow one to improve the momentum resolution. The systematic error in the determination of the polar angle was substantially decreased when information from the Z-chamber cathode was used for calibration.
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Serial Year :
1996
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Record number :
2173529
Link To Document :
بازگشت