• Title of article

    Testing and calibration of the BGO endcap calorimeter with phototriode readout for the CMD-2 detector

  • Author/Authors

    Akhmetshin، نويسنده , , R.R and Grigoriev، نويسنده , , D.N and Kazanin، نويسنده , , V.F and Logashenko، نويسنده , , I.B and Smakhtin، نويسنده , , V.P and Yudin، نويسنده , , Yu.V، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 1996
  • Pages
    2
  • From page
    509
  • To page
    510
  • Abstract
    Testing and calibration of the endcap calorimeter of the CMD-2 detector for the VEPP-2M collider is described. The endcap calorimeter consists of 680 BGO crystals with phototriode readout and works in a 1 T magnetic field. All crystals, phototriodes and preamplifiers were tested before installation. Electronic and cosmic rays calibration are regularly performed. As the calibration by cosmic rays reveals, the average energy equivalent of noise is about 1 MeV, in accordance with the design value. Long term stability of the calorimeter has been studied.
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Serial Year
    1996
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Record number

    2173582