Title of article :
Improved performance lead iodide nuclear radiation detectors
Author/Authors :
James Deich، نويسنده , , V and Roth، نويسنده , , M، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1996
Pages :
4
From page :
169
To page :
172
Abstract :
Melt-grown lead iodide single crystals with an improved transport of positive charge carriers (holes) have been developed. It is shown that an energy resolution of about 3% can be obtained for gamma-radiation at 60 keV (241Am isotope source). Improvement of the spectrometric performance of lead iodide detectors at lower gamma-ray energies has been achieved as well due to better purification of the starting material and optimization of the etching procedure. Correlation between the detector performance and the structure of lead iodide crystals has been studied using X-ray diffraction analysis.
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Serial Year :
1996
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Record number :
2173698
Link To Document :