• Title of article

    Improved performance lead iodide nuclear radiation detectors

  • Author/Authors

    James Deich، نويسنده , , V and Roth، نويسنده , , M، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 1996
  • Pages
    4
  • From page
    169
  • To page
    172
  • Abstract
    Melt-grown lead iodide single crystals with an improved transport of positive charge carriers (holes) have been developed. It is shown that an energy resolution of about 3% can be obtained for gamma-radiation at 60 keV (241Am isotope source). Improvement of the spectrometric performance of lead iodide detectors at lower gamma-ray energies has been achieved as well due to better purification of the starting material and optimization of the etching procedure. Correlation between the detector performance and the structure of lead iodide crystals has been studied using X-ray diffraction analysis.
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Serial Year
    1996
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Record number

    2173698