• Title of article

    Proton energy measurements using a NaI(T1) scintillator

  • Author/Authors

    Schneider، نويسنده , , Uwe and Renker، نويسنده , , Dieter، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 1997
  • Pages
    5
  • From page
    199
  • To page
    203
  • Abstract
    Energy measurements of protons using a NaI(T1) scintillator for energies around 100 MeV and higher, have the disadvantage that the response of the crystal has a non-linear light output versus energy relation. An experimental calibration of the light output of the crystal into energy would require a number of different monoenergetic proton beams. We have developed a method for calibrating the non-linear light output into energy using relative measurements of the energy. By applying this relative method, it is also possible to determine the absolute initial energy of the protons before they hit the experimental set-up.
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Serial Year
    1997
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Record number

    2175129