• Title of article

    Calibration of a liquid xenon gamma ray detector for the study of radiative pion decay

  • Author/Authors

    Bressi، نويسنده , , G and Carugno، نويسنده , , G and Cerdonio، نويسنده , , S and Conti، نويسنده , , E and Meneguzzo، نويسنده , , A.T and Zanello، نويسنده , , D، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 1997
  • Pages
    9
  • From page
    67
  • To page
    75
  • Abstract
    The performance of a liquid xenon time projection chamber as a detector for gamma rays between 0 and 30 MeV has been studied by comparing data collected with radioactive sources to a Monte Carlo simulation. The detector was designed for the study of the pion radiative decay. he scintillation light and the ionization charge signals were exploited. ght signals were used in the trigger logic and for the measurement of the chamber trigger acceptance. The minimum triggerable energy turned out to be 170 keV. arge signals were used to measure the gamma ray energy. The noise introduced by the electronics was about 140 keV and the minimum measurable energy 230 keV with an electric drift field of 0.5 kV/cm. e charge and light spectra were in good agreement with the simulation. erall apparatus acceptance, taking into account geometry, trigger efficiency and offline reconstruction losses, was independent of the gamma ray energy above 2 MeV and equal to 10%.
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Serial Year
    1997
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Record number

    2176323