Title of article
Preparation and standardization of large-area sources for calibration of alpha-particle contamination monitors
Author/Authors
Garcيa-Toraٌo، نويسنده , , E and Goٌi، نويسنده , , J and Gascَn، نويسنده , , J.L، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 1997
Pages
5
From page
130
To page
134
Abstract
Large-area sources of 241Am for calibration of surface-contamination monitors have been produced by incorporating the radionuclide into a thin layer of anodized Al. A measurement chamber has been developed which allows the standardization of alpha-particle emitting sources with active areas up to 12 × 17 cm2. The counter can be operated as an ionization chamber or in the proportional region. The sources can be standardized in terms of alpha-particle emission rate or activity with uncertainties in the order of 1%.
Journal title
Nuclear Instruments and Methods in Physics Research Section A
Serial Year
1997
Journal title
Nuclear Instruments and Methods in Physics Research Section A
Record number
2176361
Link To Document