Title of article :
Electrode thickness measurement of a Si(Li) detector for the SIXA array
Author/Authors :
Tikkanen، نويسنده , , T. and Hنmنlنinen، نويسنده , , K. Alanko-Huotari، نويسنده , , S.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1998
Abstract :
Cathode electrodes of the Si(Li) detector elements of the SIXA X-ray spectrometer array are formed by gold-palladium alloy contact layers. The equivalent thickness of gold in one element was measured by observing the characteristic L-shell X-rays of gold excited by monochromatised synchrotron radiation with photon energies above the L3 absorption edge of gold. The results obtained at four different photon energies below the L2 edge yield an average value of 22.4 ± 3.5 nm which is consistent with the earlier result extracted from detection efficiency measurements.
Keywords :
Gold electrodes , Synchrotron radiation , Si(Li) detectors , X-ray spectrometers , X-ray fluorescence , detector calibration