Title of article :
Calibration of the well-type germanium gamma-ray detector employing two gamma-ray spectra
Author/Authors :
Blaauw، نويسنده , , M.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1998
Pages :
8
From page :
146
To page :
153
Abstract :
Since 1992, a method has been under development at IRI to characterise a well-type gamma-ray detector by determination of a peak efficiency and a peak-to-total efficiency curve through measurement of a radionuclide emitting coincident photons, and to compute radionuclide specific counting efficiencies from those curves for “real” peaks as well as for “sum” peaks. Recent improvements to this method have indicated that now, a well-type can be as easy to calibrate and operate as a coaxial detector, once the software is in place. In this paper, the improved calibration method will be described and examples of results given.
Keywords :
?-ray spectrometry , WELL detectors , Calibration , INAA-nuclides , True coincidenec summing
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Serial Year :
1998
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Record number :
2180120
Link To Document :
بازگشت