Title of article
Development and prospects of the new gaseous detector “Micromegas”
Author/Authors
Giomataris، نويسنده , , Y.، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 1998
Pages
12
From page
239
To page
250
Abstract
We report results obtained with the novel gaseous Micromegas detector (MICRO MEsh GAseous Structure), which is under development at Saclay. A simple theory to explain the advantage of the small amplification gap (50–100 μm) is developed. A set of large detectors was exposed during several months in high-intensity particle beams. Full efficiency and a large plateau has been obtained with a 3 mm conversion gap. With a convertion gap as small as 1 mm the efficiency reaches 96%. A spatial resolution better than 60 μm has been observed with anode strips of 317 μm pitch. Simulations show that with a pitch of 100 μm and the appropriate gas an accuracy of 10 μm and a time resolution of 1 ns is within reach. This development leads to a new generation of cheap position sensitive detectors which would permit high precision tracking or vertexing close to the interaction region, in very high-rate environments.
Journal title
Nuclear Instruments and Methods in Physics Research Section A
Serial Year
1998
Journal title
Nuclear Instruments and Methods in Physics Research Section A
Record number
2180144
Link To Document