Title of article :
X-ray polarimetry with a conventional gas proportional counter through rise-time analysis
Author/Authors :
Hayashida، نويسنده , , Kiyoshi and Miura، نويسنده , , Noriyuki and Tsunemi، نويسنده , , Hiroshi and Torii، نويسنده , , Kenichi and Murakami، نويسنده , , Hiroyuki and Ohno، نويسنده , , Yoshiaki and Tamura، نويسنده , , Keisuke، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1999
Pages :
8
From page :
241
To page :
248
Abstract :
We have performed an experiment on the signal rise time of a Xe gas proportional counter using a polarized X-ray beam of synchrotron orbital radiation with energies from 10 to 40 keV. When the counter anode is perpendicular to the electric vector of the incident X-ray photons, the average rise time becomes significantly longer than that for the parallel case. This indicates that the conventional gas proportional counters are useful for X-ray polarimetry. The moderate modulation contrast of this rise-time polarimeter (M=0.1 for 10 keV X-rays and M=0.35 for 40 keV X-rays), with capability of the simultaneous measuring X-ray energies and the timing, would be useful for applications in X-ray astronomy and in other fields.
Keywords :
X-ray polarimeter , X-ray Astronomy , Synchrotron radiation , Proportional counter
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Serial Year :
1999
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Record number :
2180302