Title of article :
An electrostatic radon detector designed for water radioactivity measurements
Author/Authors :
Wang، نويسنده , , Jian-Xiong and Andersen، نويسنده , , Tom C. and Simpson، نويسنده , , John J.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1999
Pages :
9
From page :
601
To page :
609
Abstract :
The Sudbury Neutrino Observatory contains 1000 t of ultra-pure heavy water and 7000 t of very pure light water. In order to achieve the goal of determining accurately the total number of neutrinos emitted by the sun regardless of type, the radioactivity in the water must be maintained with a radiopurity of less than 10−14 g/g of 232Th and 238U in secular equilibrium with their daughters. This paper describes the design and application of an electrostatic radon detector which determines the amount of radon emanating from a column containing MnO2 used to extract radium from the water. This electrostatic radon detector has a total efficiency of 35% for detecting 222Rn and 22% for 220Rn.
Keywords :
Electrostatic radon detector , Solar neutrino , Water purification , Alpha particle spectrometry , Radioactivity assay
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Serial Year :
1999
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Record number :
2180339
Link To Document :
بازگشت