Title of article
Neutron proportional counter design for high gamma-ray environments
Author/Authors
Beddingfield، نويسنده , , D.H. and Menlove، نويسنده , , H.O and Johnson، نويسنده , , N.H، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 1999
Pages
6
From page
35
To page
40
Abstract
A parametric analysis of the performance of 3He neutron proportional detector tubes in mixed gamma-ray/neutron environments has been performed. The objective of this study was to determine the optimum tube design configuration for minimized gamma-ray sensitivity based upon commercially available components. The parameters examined in the study were the tube wall material, the admix gas identity, the total fill pressure, and the tube lining material. The study resulted in the quantification of the limitations of typical 3He tube designs and in the development of a new tube design which extends the usefulness of 3He tubes in mixed gamma-ray/neutron environments.
Keywords
3He neutron proportional counter , Gamma-ray interactions , Gamma-ray pile-up
Journal title
Nuclear Instruments and Methods in Physics Research Section A
Serial Year
1999
Journal title
Nuclear Instruments and Methods in Physics Research Section A
Record number
2180350
Link To Document