Title of article :
Signal formation in a CdZnTe imaging detector with coplanar pixel and control electrodes
Author/Authors :
Mayer، نويسنده , , M and Hamel، نويسنده , , L.A and Tousignant، نويسنده , , O and Macri، نويسنده , , J.R and Ryan، نويسنده , , J.M and McConnell، نويسنده , , M.L and Jordanov، نويسنده , , V.T and Butler، نويسنده , , J.F and Lingren، نويسنده , , C.L، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1999
Pages :
5
From page :
190
To page :
194
Abstract :
In order to gain insight on a proposed CZT imaging device with orthogonal coplanar anodes currently under development, measurements have been made on an available commercial device with pixels and control electrodes. Charge signals induced on both the pixel and control electrodes by alphas incident at two positions on the cathode side are measured and compared with model predictions. Signals produced by γ-rays are also discussed. Measurements and simulations are found to be in reasonably good agreement. It is concluded that the proposed anode geometry might exhibit excellent energy resolution and good position resolution in two dimensions as well as some position information along the depth dimension.
Keywords :
CdznTe imaging detector , Coplaner pixel , Control electrodes
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Serial Year :
1999
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Record number :
2180380
Link To Document :
بازگشت