• Title of article

    Two-dimensional position sensitive silicon photodiode as a charged particle detector

  • Author/Authors

    Kova?evi?، نويسنده , , K and Zadro، نويسنده , , M، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 1999
  • Pages
    5
  • From page
    103
  • To page
    107
  • Abstract
    A two-dimensional position sensitive silicon photodiode has been tested for measurement of position and energy of charged particles. Position nonlinearity and resolution, as well as energy resolution and ballistic deficit were measured for 5.486 MeV α-particles. The results obtained for different pulse shaping time constants are presented.
  • Keywords
    photodiode , Position sensitive detector , Nonlinearity , Ballistic deficit , RESOLUTION
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Serial Year
    1999
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Record number

    2180541