Title of article :
An achromatic multipassage magnetic mass spectrometer
Author/Authors :
Boulanger، نويسنده , , P and Baril، نويسنده , , M، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1999
Pages :
5
From page :
136
To page :
140
Abstract :
A design providing achromatic correction to a multipassage magnetic mass spectrometer previously described by the author is presented. The energy spatial dispersion caused by repeated passages in the magnetic prism is corrected by three supplementary mirrors placed in a reinjection loop. From this study one can see that we simultaneously eliminate the energy dispersion term CΔE/E and the opening angle aberration term Cα2 and we may also eliminate the coupled aberration term CαΔE/E.
Keywords :
Magnetic mass spectrometer , Mass spectrometer , Multipassage mass spectrometer , Achromatic mass spectrometer
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Serial Year :
1999
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Record number :
2181180
Link To Document :
بازگشت