Title of article :
A cryogenic current-measuring device with nano-ampere resolution at the storage ring TARN II
Author/Authors :
Tanabe، نويسنده , , T. and Chida، نويسنده , , K. and Shinada، نويسنده , , K.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1999
Pages :
10
From page :
455
To page :
464
Abstract :
In cooler-ring experiments, an accurate and non-destructive current measurement is essential for determining the reaction cross sections. The lowest current which can be measured by the DC current transformer commonly used so far is some μA. In order to measure a low-beam current from nA to μA, we made a cryogenic current-measuring device using a superconducting quantum interference devices (SQUID), and measured the circulating ion current at the cooler ring TARN II. This paper gives the design and performance of the device.