Title of article :
Model-based pulse shape correction for CdTe detectors
Author/Authors :
Bargholtz، نويسنده , , Chr and Fumero، نويسنده , , E and Mهrtensson، نويسنده , , L، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1999
Abstract :
We present a systematic method to improve energy resolution of CdTe-detector systems with full control of the efficiency. Sampled pulses and multiple amplifier data are fitted by a model of the pulse shape including the deposited energy and the interaction point within the detector as parameters. We show the decisive improvements of spectral resolution and photo-peak efficiency that is obtained without distortion of spectral shape. The information concerning the interaction depth of individual events can be used to discriminate between beta particles and gamma quanta.
Keywords :
CdTe detector , Pulse-shape analysis , Gamma-ray spectrometry
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Journal title :
Nuclear Instruments and Methods in Physics Research Section A