• Title of article

    Performance of superconducting transition edge bolometers for infrared photometry and X-ray micro-calorimetry

  • Author/Authors

    Hoevers، نويسنده , , H.F.C and Bento، نويسنده , , A.C and Bruijn، نويسنده , , M.P and Frericks، نويسنده , , M and Kiewiet، نويسنده , , F.B and Mels، نويسنده , , W.A and de Korte، نويسنده , , P.A.J، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 1999
  • Pages
    5
  • From page
    247
  • To page
    251
  • Abstract
    The performance of voltage biased superconducting transition edge bolometers for infrared photometry and X-ray micro-calorimetry is described. Experimental results for the resistance–temperature curve of the superconducting phase thermometer, the current–voltage characteristic of the bolometer, and its current noise are presented and evaluated. Excess current noise turns out to be present when the bolometers are biased in the superconducting transition. The possible origin of this excess noise is shortly discussed. For the infrared bolometer an NEP of 2 ×10−17 W/Hz at 270 mK is obtained. The X-ray micro-calorimeter resolution is 60 eVFWHM at 155 mK for 5.9 keV X-rays, a factor 2.6 above the thermodynamical limit.
  • Keywords
    electrothermal feedback , Infrared detection , x-ray detection , Superconducting transition edge bolometer
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Serial Year
    1999
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Record number