Title of article :
Spectroscopic measurements with a silicon drift detector having a continuous implanted drift cathode-voltage divider
Author/Authors :
Bonvicini، نويسنده , , V. and Bürger، نويسنده , , P. and DʹAcunto، نويسنده , , L. and Franck، نويسنده , , D. Di Gregorio، نويسنده , , A. and Pihet، نويسنده , , P. and Rashevsky، نويسنده , , A. and Vacchi، نويسنده , , A. and Vinogradov، نويسنده , , L. and Zampa، نويسنده , , N.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2000
Pages :
5
From page :
471
To page :
475
Abstract :
A silicon drift detector (SDD) prototype where the drift electrode also plays the role of a high-voltage divider has been realised and characterised for spectroscopic applications at near-room temperatures. Among the advantages of this design, is the absence of metal on the sensitive surface which makes this detector interesting for soft X-rays. The detector prototype has a large sensitive area (2×130 mm2) and the charge is collected by two anodes (butterfly-like detector). ergy resolution of a such a detector has been investigated at near-room temperatures using a commercial, hybrid, low-noise charge-sensitive preamplifier. The results obtained for the X-ray lines from 55Fe and 241Am are presented.
Keywords :
Silicon drift detectors , X-ray spectroscopy
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Serial Year :
2000
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Record number :
2184427
Link To Document :
بازگشت