    Simulation and experimental results for the detection of conversion electrons with gas proportional scintillation counters

  • Author/Authors

    Rachinhas، نويسنده , , P.J.B.M. and Simُes، نويسنده , , P.C.P.S. and Lopes، نويسنده , , J.A.M. and Dias، نويسنده , , T.H.V.T. and Morgado، نويسنده , , R.E. and dos Santos، نويسنده , , J.M.F. and Stauffer، نويسنده , , A.D. and Conde، نويسنده , , C.A.N.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2000
  • Pages
    11
  • From page
    468
  • To page
    478
  • Abstract
    The application of gas proportional scintillation counters (GPSC) to the detection and identification of conversion electrons in the medium energy range is investigated. Experimental and Monte Carlo results are presented for the response of a xenon GPSC, filled at atmospheric pressure, to the decay of a 109Cd source. This source emits 88.0 keV γ-rays, eL=84.6 keV and eK=62.5 keV conversion electrons, as well as fluorescence X-rays and Auger electrons. Good agreement is found between the measured and the calculated energy spectra. The response to higher-energy electrons is investigated by Monte Carlo simulation, by considering a hypothetical GPSC filled with xenon at 10 atm and doped with the 133mXe metastable isotope. The calculated energy spectra for the absorption of the 133mXe 233.2 keV γ-rays, eK=198.6 keV and eL=228.4 keV conversion electrons, as well as fluorescence X-rays and Auger electrons, are presented and discussed.
  • Keywords
    Conversion electrons , Gas scintillation detectors , Monte Carlo simulation , Digital pulse processing
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Serial Year
    2000
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Record number