Title of article
Fiske modes in superconducting tunnel junction detectors
Author/Authors
Friedrich، نويسنده , , S and Cunningham، نويسنده , , M.F and Frank، نويسنده , , M and Labov، نويسنده , , S.E and Barfknecht، نويسنده , , A.T and Cramer، نويسنده , , S.P، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2000
Pages
5
From page
151
To page
155
Abstract
We are developing superconducting tunnel junction (STJ) X-ray spectrometers with high-energy resolution (<10 eV for keV X-rays) and high count rate capabilities (>10 000 counts/s). We have examined the influence of Fiske mode resonances on energy resolution, maximum count rate and area of Nb–Al–AlOx–Al–Nb STJ detectors. At low count rate, these detectors have an energy resolution of 2.3 eV FWHM at 70 eV. At 20 000 counts/s, the resolution is 8.2 eV FWHM at 277 eV. We observe the optimum detector performance when the influence of Fiske modes is minimized.
Keywords
Cryogenic X-ray detectors , Fiske modes , Superconducting tunnel junctions
Journal title
Nuclear Instruments and Methods in Physics Research Section A
Serial Year
2000
Journal title
Nuclear Instruments and Methods in Physics Research Section A
Record number
2185608