Author/Authors :
Bravin، نويسنده , , G and Bruckmayer، نويسنده , , M and Frank، نويسنده , , T and Giordano، نويسنده , , S and Loidl، نويسنده , , M and Meier، نويسنده , , O and Meunier، نويسنده , , P and Pergolesi، نويسنده , , D and Prِbst، نويسنده , , F and Seidel، نويسنده , , W and Sisti، نويسنده , , M and di Stefano، نويسنده , , P and Stodolsky، نويسنده , , L and Uchaikin، نويسنده , , S and Zerle، نويسنده , , L، نويسنده ,
Abstract :
We report on simultaneous detection of phonons and scintillation light using a cryogenic calorimeter composed of a scintillating CaWO4 crystal and a separate light detector, both instrumented with a tungsten superconducting phase transition thermometer operating at about 12 mK. Clear discrimination between electron and nuclear recoils was demonstrated by irradiating the detector with electrons, photons and neutrons. The shape of the nuclear recoil energy spectrum induced by the neutrons could be well reproduced by simulations.