Title of article :
Factors affecting the performance of detectors that use secondary electron emission from a thin foil to determine ion impact position
Author/Authors :
Shapira، نويسنده , , D and Lewis، نويسنده , , T.A and Hulett Jr.، نويسنده , , L.D and Ciao، نويسنده , , Z، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2000
Pages :
12
From page :
396
To page :
407
Abstract :
The factors affecting the performance of a Position-Sensitive Timing Detector (PSTD), based on the detection of secondary electrons ejected from a thin foil by the passing ion, were investigated. Different contributions to the observed (generally poor) position resolution were investigated. The effects of multiple scattering in the foil were determined first. We then measured the contribution of the position-sensitive detector and signal-handling electronics. Finally, the effect of electron transport, from the point of emission at the foil surface to the micro-channel plate detector (MCP) surface, was measured and evaluated. Our measurements, as well as a simulation of the electron transport, lead us to conclude that for detectors with foil-to-MCP distance ≥5 cm, electron transport limits the devices currently in use to resolution worse than 2 mm (FWHM).
Keywords :
Limits to time and position resolution , Micro channel plate detectors , Timing and position measurement , Ions and nuclei , Electron emission from surface , Electron acceleration and detection
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Serial Year :
2000
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Record number :
2186998
Link To Document :
بازگشت