Title of article
Performance of ACCOS, an Automatic Crystal quality Control System for the PWO crystals of the CMS calorimeter
Author/Authors
Auffray، نويسنده , , E and Chevenier، نويسنده , , G and Freire، نويسنده , , M and Lecoq، نويسنده , , P and Le Goff، نويسنده , , J.M and Marcos، نويسنده , , R and Drobychev، نويسنده , , G and Missévitch، نويسنده , , O and Oskine، نويسنده , , A and Zouevsky، نويسنده , , R and Peigneux، نويسنده , , J.P and Schneegans، نويسنده , , M، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2001
Pages
17
From page
325
To page
341
Abstract
Nearly 80 000 PWO crystals for the CMS electromagnetic calorimeter will arrive at CERN/Geneva and INFN-ENEA/Rome between now and year 2004. The stringent specifications on their dimensions and optical quality have to be verified prior to their formal acceptation. Automatic systems for measuring the critical parameters of each crystal and recording them in a database have been designed and constructed. The first machine is now in stable operation at CERN. In this note, the performance of each instrument, based on the measurements on ∼1000 pre-production crystals, is analysed in terms of stability and compared to the results of conventional benches.
Journal title
Nuclear Instruments and Methods in Physics Research Section A
Serial Year
2001
Journal title
Nuclear Instruments and Methods in Physics Research Section A
Record number
2188507
Link To Document