Title of article :
The bio-crystallography beamline (BL41XU) at SPring-8
Author/Authors :
Kawamoto، نويسنده , , Masahide and Kawano، نويسنده , , Yoshiaki and Kamiya، نويسنده , , Nobuo، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2001
Pages :
5
From page :
1375
To page :
1379
Abstract :
The bio-crystallography beamline (BL41XU), one of two pilot beamlines at SPring-8, was constructed using a standard in-vacuum-type undulator and opened for general users from domestic and overseas countries. Many tests and improvements were carried out on beamline elements and equipment for macromolecular crystallography, especially on the so-called “pin-post” water cooling crystal of rotated-inclined double crystal monochromator. The maximum brilliance at sample position reached to 4×1015 photons/s/mm2/mrad2 at an X-ray energy of 11 keV. Commercially available X-ray detectors of CCD and imaging plate were installed in the experimental station. A beamline control software system for beam tracking and an on-line reader for large-format imaging plate were newly developed.
Keywords :
macromolecular crystallography , High brilliance , In-vacuum-type undulator , Wide energy range , Pin-post water cooling monochromator , SPring-8
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Serial Year :
2001
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Record number :
2192279
Link To Document :
بازگشت