Title of article :
X-ray absorption spectroscopy of diluted system by undulator photon source and multi-element solid-state detector
Author/Authors :
Tanida، نويسنده , , Hajime and Ishii، نويسنده , , Masashi، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2001
Abstract :
In order to measure the extended X-ray absorption fine structure (EXAFS) spectrum of an ultra-diluted system, an optics and detector control system for a synchrotron radiation beamline is developed. The undulator gap width is continuously tuned to obtain the maximum X-ray photon flux during the energy scan for the EXAFS measurement. A piezoelectric translator optimizes the parallelism of the double crystal in a monochromator at each measurement point to compensate for mechanical errors of the monochromator, resulting in a smooth and intense X-ray photon flux during the measurement. For a detection of a weak fluorescence signal from diluted samples, a 19-element solid-state detector and digital signal processor are used. A K-edge EXAFS spectrum of iron in a myoglobin aqueous solution with a concentration of 5.58 parts per million was obtained by this system.
Keywords :
Ultra diluted system , 19-element solid-state detector , Undulator gap scan , digital signal processor
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Journal title :
Nuclear Instruments and Methods in Physics Research Section A