• Title of article

    Neutron light output response and resolution functions in EJ-309 liquid scintillation detectors

  • Author/Authors

    Enqvist، نويسنده , , Andreas and Lawrence، نويسنده , , Christopher C. and Wieger، نويسنده , , Brian M. and Pozzi، نويسنده , , Sara A. and Massey، نويسنده , , Thomas N.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2013
  • Pages
    8
  • From page
    79
  • To page
    86
  • Abstract
    Neutron light output response functions and detector resolution functions were measured at Ohio Universityʹs tandem Van de Graaff generator for three cylindrical EJ-309 liquid scintillator cells, having dimensions 12.7 ( ∅ )-by-12.7, 7.6-by-7.6, and 7.6-by-5.1 cm. A 7.44 MeV deuteron beam was used on an 27Al target generating a continuous spectrum over the energy range from a few hundred keV to over 10 MeV. The light output response functions are determined using an exponential fit. Detector resolution functions are obtained for the 12.7-by-12.7 and 7.6-by-7.6 cm detectors. It is demonstrated that the dependence on detector size is important for the light output response functions, but not to the same extent for the resolution function, even when photomultiplier tubes, detector material, and other detector characteristics are carefully matched.
  • Keywords
    Liquid scintillation detectors , Pulse shape discrimination , Detector characterization , neutrons , Time-of-flight
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Serial Year
    2013
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Record number

    2194038