Title of article :
Impact of digitization for timing and pulse shape analysis of scintillator detector signals
Author/Authors :
Hellesen، نويسنده , , C. and Skiba، نويسنده , , M. and Ericsson، نويسنده , , G. and Andersson Sundén، نويسنده , , Dale E. and Binda، نويسنده , , F. and Conroy، نويسنده , , S. and Eriksson، نويسنده , , J. and Weiszflog، نويسنده , , M.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2013
Pages :
6
From page :
135
To page :
140
Abstract :
In this paper we investigate the effects of full digitization of scintillator signals. The requirements on the analog to digital converter (ADC), in terms of sampling rate (fs) and bit resolution, are investigated. Two applications for scintillator detectors are studied, pulse timing and particle species identification. We find that signal reconstruction using sinc interpolation can be used e.g. for high-precision timing of a sampled electric pulse. Timing performances better than 6 ps (FWHM) were obtained if fs equals or exceeds twice the maximum frequency of the scintillator pulse. Failing to meet this criterion deteriorates both the performance of pulse timing and particle identification. We find that the bit resolution of the ADC is very important also for timing of pulses.
Keywords :
Pulse shape analysis , Digitization , Pulse timing , scintillator
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Serial Year :
2013
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Record number :
2194399
Link To Document :
بازگشت