• Title of article

    Improving the effectiveness of a low-energy Compton suppression system

  • Author/Authors

    Britton، نويسنده , , R. and Burnett، نويسنده , , J.L. and Davies، نويسنده , , A.V. and Regan، نويسنده , , P.H.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2013
  • Pages
    5
  • From page
    64
  • To page
    68
  • Abstract
    A novel method for collecting and processing coincidence data from a Compton Suppressed Low Energy Photon Spectrometer (LEPS) is presented, greatly simplifying the current setup and extending the suppression abilities of the system. Offline analysis is used, eliminating the need to discard coincidence data when vetoing coincident events with fast-timing electronics. Additional coincident events are identified that are usually missed, and which represent interactions in the active NaI(Tl) shield prior to an interaction in the LEPS detector. By suppressing these events, the Compton Suppression factor was improved by 144% for the 661.66 keV decay line in a 137Cs source. The geometry used for this particular Compton suppression system is highly sensitive to these effects, however similar event profiles are expected in all coincidence systems.
  • Keywords
    gamma spectroscopy , Environmental monitoring , Compton suppression , coincidence
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Serial Year
    2013
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Record number

    2194633