• Title of article

    3D position readout from thick scintillators

  • Author/Authors

    Antich، نويسنده , , P and Malakhov، نويسنده , , N and Parkey، نويسنده , , R and Slavin، نويسنده , , N and Tsyganov، نويسنده , , E، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2002
  • Pages
    6
  • From page
    782
  • To page
    787
  • Abstract
    A novel technique has been developed and tested for the three-dimensional measurement of position in SPECT-PET detectors. Results are presented for 2 and 20 mm thick NaI(Tl) planar crystals. In a plane of crystal, a coordinate resolution of about 1 mm (rms error) is demonstrated. The depth of interaction (DOI) is measured with an rms error of about 2 mm using light cone parameters.
  • Keywords
    Scintillation detectors , Position resolution , Depth of interaction
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Serial Year
    2002
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Record number