    Basic performance of a position-sensitive tissue-equivalent proportional chamber (PS-TEPC)

  • Author/Authors

    Kishimoto، نويسنده , , Yuji C. Sasaki، نويسنده , , Shinichi and Saito، نويسنده , , Kiwamu and Takahashi، نويسنده , , Kazutoshi and Doke، نويسنده , , Tadayoshi and Miuchi، نويسنده , , Kentaro and Fuse، نويسنده , , Tetsuhito and Nagamatsu، نويسنده , , Aiko and Uchihori، نويسنده , , Yukio and Kitamura، نويسنده , , Hisashi and Terasawa، نويسنده , , Kazuhiro، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2013
  • Pages
    4
  • From page
    591
  • To page
    594
  • Abstract
    The characteristics and performance of the prototype of a Position-Sensitive Tissue-Equivalent Proportional Chamber (PS-TEPC), consisting of a miniaturized micro-pixel chamber ( μ - PIC ), were investigated. The gas–gain curve of the μ - PIC and an approximate function with which to convert measured collected charge into the primary charge ionized by incident heavy-ion particles were obtained. The measured Linear Energy Transfer (LET) indicated good agreement with that of the injected beam calculated using SRIM2008 code in the LET range from 0.4 to 180 keV / μ m , which roughly covers the whole LET range required for space dosimeters. As expected, the measured energy resolution improved with increasing LET of the injected beam. It was estimated to be 27% for 0.4 keV / μ m , which is just about the LET of minimum-ionizing particles. Consequently, it was confirmed that the PS-TEPC exhibits no significant undesirable feature but is well suited for space dosimetry.
  • Keywords
    Space dosimeter , LET , PS-TEPC
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Serial Year
    2013
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Record number

    2195238