Title of article :
Characteristics of a normoxic polymethacrylic acid gel dosimeter for a 72-MeV proton beam
Author/Authors :
Bong، نويسنده , , Jihye and Shin، نويسنده , , Dongho and Kwon، نويسنده , , Soo-Il، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2014
Abstract :
The characteristics of a normoxic polymethacrylic acid gel dosimeter for a 72-MeV proton beam were evaluated. A polymer gel dosimeter was synthesized using gelatin, methacrylic acid, hydroquinone, tetrakis(hydroxymethyl) phosphonium chloride, and highly purified distilled water. The dosimeter was manufactured by placement in a polyethylene (PE) container. Irradiated dosimeters were analyzed to determine the transverse relaxation time (T2) using a 1.5-T MRI. A calibration curve was obtained as a function of the absorbed dose.
g curve made by irradiating the gel with mono-energy was compared with the results for a parallel plate ionization chamber. The spread-out Bragg peak (SOBP) range and distal dose fall-off (DDF) were comparatively analyzed by comparing the irradiated gel with a spread-out Bragg peak against with the ion chamber.
, the gelʹs usefulness as a dosimeter for therapeutic radiation quality assurance was evaluated by obtaining its practical field size, flatness, and symmetry, through comparison of the profiles of the gel and ion chamber.
Keywords :
proton , Nomoxic polymethacrylic acid gel , dosimetry , quenching
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Journal title :
Nuclear Instruments and Methods in Physics Research Section A