Title of article :
Characterization of a cubic EJ-309 liquid scintillator detector
Author/Authors :
Tomanin، نويسنده , , A. and Paepen، نويسنده , , J. and Schillebeeckx، نويسنده , , P. and Wynants، نويسنده , , R. and Nolte، نويسنده , , R. and Lavietes، نويسنده , , A.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2014
Abstract :
A cubic EJ-309 liquid scintillator of 10 cm width has been characterized for its response to γ-rays and neutrons. Response functions to γ-rays were measured with calibrated radionuclide γ-ray sources in the energy range from 400 keV to 6 MeV. Response functions for neutrons were obtained from measurements at the PTB Van de Graaff accelerator with quasi-monoenergetic neutron beams in the energy range from 500 keV to 2.7 MeV, and at the PTB cyclotron with time-of-flight (TOF) measurements in the energy range from 2.5 to 14 MeV. The light output and resolution functions for electrons and protons were derived by a least squares adjustment to experimental data using theoretical response functions determined with Monte Carlo simulations. The simulated response function for neutron was validated by results of measurements with an AmBe neutron source which was characterized for its total neutron intensity. The results indicate that the cubic EJ-309 detector is suitable for use in mixed γ-ray and neutron fields.
Keywords :
Pulse shape discrimination (PSD) , EJ-309 , neutron detector , Time-of-flight (TOF) , Liquid scintillator , Response functions
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Journal title :
Nuclear Instruments and Methods in Physics Research Section A