Harkness-Brennan، نويسنده , , L.J. and Judson، نويسنده , , D.S. and Boston، نويسنده , , A.J. and Boston، نويسنده , , H.C. and Colosimo، نويسنده , , S.J. and Cresswell، نويسنده , , J.R. and Nolan، نويسنده , , P.J. and Adekola، نويسنده , , A.S. and Colaresi، نويسنده , , J. and Cocks، نويسنده , , J.F.C. and Mueller، نويسنده , , W.F.، نويسنده ,
Abstract :
The spectroscopic and charge collection performance of a BE2825 Broad Energy Germanium (BEGe) detector has been experimentally investigated. The efficiency and energy resolution of the detector have been measured as a function of energy and the noise contributions to the preamplifier signal have been determined. Collimated gamma-ray sources mounted on an automated 3-axis scanning table have been used to study the variation in preamplifier signal shape with gamma-ray interaction position in the detector, so that the position-dependent charge collection process could be characterised. A suite of experimental measurements have also been undertaken to investigate the performance of the detector as a function of bias voltage and we report on anomalous behaviour observed when the detector was operating close to the depletion voltage.