Title of article
Evaluation of two thermal neutron detection units consisting of ZnS/6LiF scintillating layers with embedded WLS fibers read out with a SiPM
Author/Authors
Mosset، نويسنده , , J.-B. and Stoykov، نويسنده , , A. and Greuter، نويسنده , , U. and Hildebrandt، نويسنده , , M. and Schlumpf، نويسنده , , N. and Van Swygenhoven، نويسنده , , H.، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2014
Pages
6
From page
299
To page
304
Abstract
Two single channel detection units for thermal neutron detection are investigated in a neutron beam. They consist of two ZnS/6LiF scintillating layers sandwiching an array of WLS fibers. The pattern of these units can be repeated laterally and vertically in order to build up a one-dimensional position sensitive multi-channel detector with the needed sensitive surface and with the required neutron absorption probability. The originality of this work arises from the fact that the WLS fibers are read out with SiPMs instead of the traditionally used PMTs or MaPMTs. The signal processing system is based on a photon counting approach. For SiPMs with a dark count rate as high as 0.7 MHz, a trigger efficiency of 80% is achieved together with a system background rate lower than 10 − 3 Hz and a dead time of 30 μ s . No change of performance is observed for neutron count rates of up to 3.6 kHz.
Keywords
MPPC , WLS fibers , Photon counting , Thermal neutron detector , SiPM , ZnS/6LiF scintillator
Journal title
Nuclear Instruments and Methods in Physics Research Section A
Serial Year
2014
Journal title
Nuclear Instruments and Methods in Physics Research Section A
Record number
2196453
Link To Document