Title of article :
The pin pixel detector—neutron imaging
Author/Authors :
Bateman، نويسنده , , J.E and Connolly، نويسنده , , J.F and Derbyshire، نويسنده , , G.E. and Duxbury، نويسنده , , D.M and Marsh، نويسنده , , A.S and Rhodes، نويسنده , , N.J and Schooneveld، نويسنده , , E.M and Simmons، نويسنده , , J.E and Stephenson، نويسنده , , R، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2002
Pages :
10
From page :
596
To page :
605
Abstract :
The development and testing of a neutron gas pixel detector intended for application in neutron diffraction studies is reported. Using standard electrical connector pins as point anodes, the detector is based on a commercial 100 pin connector block. A prototype detector of aperture 25.4 mm×25.4 mm has been fabricated, giving a pixel size of 2.54 mm which matches well to the spatial resolution typically required in a neutron diffractometer. A 2-Dimensional resistive divide readout system has been adapted to permit the imaging properties of the detector to be explored in advance of true pixel readout electronics. The timing properties of the device match well to the requirements of the ISIS-pulsed neutron source.
Keywords :
Neutron diffraction , Pixel detector , Proportional counter , Pin anode