• Title of article

    The X-ray response of InP

  • Author/Authors

    Owens، نويسنده , , Alan and Bavdaz، نويسنده , , M and Gostilo، نويسنده , , V and Gryaznov، نويسنده , , D and Loupilov، نويسنده , , A and Peacock، نويسنده , , A and Sipila، نويسنده , , H، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2002
  • Pages
    6
  • From page
    435
  • To page
    440
  • Abstract
    We present the results of X-ray measurements on a prototype InP detector. The device was fabricated from Fe-doped bulk material of size 3×3×0.18 mm3. X-ray measurements have been carried out using a number of radioactive and fluorescent target sources. The detector energy response function was found to be linear over the energy range 5.9–88 keV with an average rms non-linearity of 0.7%, consistent with statistics. At a detector temperature of −60°C, the FWHM energy resolution under full-area illumination was 2.5 at 5.9 keV rising to 12 at 88 keV. Analysis of the energy resolution function indicates that poor charge transport presently limits the performance of InP detectors.
  • Keywords
    Compound semiconductors , InP , X-rays
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Serial Year
    2002
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Record number

    2196951