Title of article :
A new positron lifetime spectrometer using a fast digital oscilloscope and BaF2 scintillators
Author/Authors :
Saito، نويسنده , , H and Nagashima، نويسنده , , Y and Kurihara، نويسنده , , T and Hyodo، نويسنده , , T، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2002
Abstract :
A new positron lifetime spectrometer has been constructed and tested. It consists of a state-of-the-art digital oscilloscope (LeCroy Wavepro 960) and scintillation detectors based on Hamamatsu H3378 (R2083Q) photomultiplier tubes and BaF2 scintillators. A very good time resolution is obtained with scintillators of commonly used sizes (28 mm in diameter, 10–20 mm thick) and wide energy window settings (40–50%). The time resolution of 144 ps full-width at half-maximum (FWHM) is achieved for the positron lifetime measurement, in which time differences are measured between the arrivals of a 1.27 MeV γ-ray from 22Na and one of the two positron annihilation γ-rays of 0.511 MeV. The time resolution is further improved with a ‘double-stop’ setup, where timing information carried by the positron annihilation γ-ray pair is combined to determine the time of the annihilation. The time resolution with this setup is 119 ps FWHM. This improvement is interpreted in terms of the energy deposit in the scintillation detectors by γ-rays. Applications of the new spectrometer to other scintillation timing measurements are also described.
Keywords :
Digital , Positron annihilation , ?-Ray , scintillation detector , Lifetime spectrometer , digital oscilloscope
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Journal title :
Nuclear Instruments and Methods in Physics Research Section A