• Title of article

    Comment on “Perspectives of medical X-ray imaging”

  • Author/Authors

    Taibi، نويسنده , , A and Baldelli، نويسنده , , P and Tuffanelli، نويسنده , , A and Gambaccini، نويسنده , , M، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2002
  • Pages
    3
  • From page
    685
  • To page
    687
  • Abstract
    In the paper “Perspectives of medical X-ray imaging” (Nucl. Instr. and Meth. A 466 (2001) 99) the authors infer, from simple approximations, that the use of HOPG monochromator has no advantage in mammography compared to existing systems. We show that in order to compare imaging properties of different X-ray sources it is necessary to evaluate the spectra after the attenuation of the tissue to be imaged. Indeed, quasi-monochromatic X-ray sources have the potential to enhance image contrast and to reduce patient dose.
  • Keywords
    mammography , Quasi-monochromatic X-rays , X-ray sources , Mosaic crystals
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Serial Year
    2002
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Record number