Title of article
Comment on “Perspectives of medical X-ray imaging”
Author/Authors
Taibi، نويسنده , , A and Baldelli، نويسنده , , P and Tuffanelli، نويسنده , , A and Gambaccini، نويسنده , , M، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2002
Pages
3
From page
685
To page
687
Abstract
In the paper “Perspectives of medical X-ray imaging” (Nucl. Instr. and Meth. A 466 (2001) 99) the authors infer, from simple approximations, that the use of HOPG monochromator has no advantage in mammography compared to existing systems. We show that in order to compare imaging properties of different X-ray sources it is necessary to evaluate the spectra after the attenuation of the tissue to be imaged. Indeed, quasi-monochromatic X-ray sources have the potential to enhance image contrast and to reduce patient dose.
Keywords
mammography , Quasi-monochromatic X-rays , X-ray sources , Mosaic crystals
Journal title
Nuclear Instruments and Methods in Physics Research Section A
Serial Year
2002
Journal title
Nuclear Instruments and Methods in Physics Research Section A
Record number
2197016