• Title of article

    Wide energy range efficiency calibration method for Ge detectors

  • Author/Authors

    Molnلr، نويسنده , , G.L. and Révay، نويسنده , , Zs. and Belgya، نويسنده , , T.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2002
  • Pages
    20
  • From page
    140
  • To page
    159
  • Abstract
    A new method is proposed for the relative efficiency calibration of HPGe detectors in the energy range from 50 keV to 11 MeV. By simultaneously fitting all separately measured data sets from calibrated and uncalibrated multigamma sources as well as neutron capture γ-rays with a sum of orthogonal polynomials on a log–log scale, accuracies better than 0.5% between 100 and 3500 keV and better than 1% up to 6000 keV could be achieved for relative full energy peak efficiencies. Dividing the energy range and fitting piecewise with lower-order polynomials does not offer any advantage. Moreover, none of the popular non-linear empirical functions are able to fit the full range. The polynomial fit is nearly indistinguishable from a semiempirical fit. If at least one calibrated γ-ray source is included, the method can be used for absolute efficiencies as well.
  • Keywords
    efficiency calibration , Multigamma source , Neutron capture gamma ray , Ge detector , Polynomial fit
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Serial Year
    2002
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Record number