Acosta، نويسنده , , D and Klimenko، نويسنده , , S and Konigsberg، نويسنده , , J and Korytov، نويسنده , , A and Mitselmakher، نويسنده , , G and Necula، نويسنده , , V and Nomerotsky، نويسنده , , A and Pronko، نويسنده , , A and Sukhanov، نويسنده , , A and Safonov، نويسنده , , A and Tsybychev، نويسنده , , D and Wang، نويسنده , , Terence S.M. and Wong، نويسنده , , M، نويسنده ,
Abstract :
We describe the initial performance of the detector used for the luminosity measurement in the CDF experiment in Run II at the Tevatron. The detector consists of low-mass gaseous Cherenkov counters with high light yield (∼100 photoelectrons) and monitors the process of inelastic pp̄ scattering. It allows for several methods of precise luminosity measurements at peak instantaneous luminosities of 2×1032 cm−2 s−1, corresponding to an average of six pp̄ interactions per bunch crossing.