Title of article
Proposal for enhanced spatial resolution and efficiency of a fast-neutron scintillator detector by use of a parabolic mirror
Author/Authors
de Haan، نويسنده , , V.O. and van der Hagen، نويسنده , , T.H.J.J.، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2003
Pages
5
From page
881
To page
885
Abstract
The limitation on the detector efficiency due to the required spatial resolution of a fast-neutron imaging system is known as the detector problem. Here, this problem is described and a proposal for a novel fast-neutron imaging system is given to overcome this problem. The novelty is to replace the standard flat mirror in an imaging system by a parabolic mirror with its focal point at the fast-neutron source position. The spatial resolution will only be limited by the intrinsic resolution of the scintillator of the order of 0.06 mm. The intrinsic fast-neutron detection efficiency can be increased to about 20–30% without loss of spatial resolution.
Keywords
Fast-neutron radiography , Detector Efficiency , Spatial resolution