Title of article :
A high count-rate spectroscopy amplifier
Author/Authors :
Ignatyev، نويسنده , , O.V and Pulin، نويسنده , , A.D، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2004
Pages :
7
From page :
160
To page :
166
Abstract :
The paper describes a fast spectroscopy amplifier designed for application in conjunction with a natural diamond detector (NDD)-based spectrometer of neutrons, high-energy atoms and charged particles in thermonuclear fusion plasma experiments. plifier has the following distinctive features: a non-conventional shaper of spectrometric pulses; a precision “digital” base-line stabilizer; an effective built-in pile-up rejecter operating in a broad band of amplitudes. plifier is designed for applications at count rates up to 106 Hz with the time-to-peak for shaped pulses Tpeak⩾220 ns, base-level pulse width Tw⩾560 ns, and the resolving time of the pile-up inspector Tres⩾45 ns.
Keywords :
Plasma experiments , diamond , “True Gaussian” pulses , DL-differentiator , Pile-up inspector