Title of article :
An automated fast transfer system at the research reactor Munich
Author/Authors :
Li، نويسنده , , X and Henkelmann، نويسنده , , R and Baumgنrtner، نويسنده , , F، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2004
Pages :
7
From page :
470
To page :
476
Abstract :
An automated fast sample transfer system for the determination of measurements of short-lived isotopes has been constructed at the Research Reactor Munich FRM. Instrumental neutron activation analysis and analysis of fissile materials by delayed neutron counting are performed on this transfer device. A LabVIEWTM program on a PC with WindowsNT operating system controls the whole pneumatic system via a PCI-1200 interface and four signal conditioning SCXI-modules of National Instruments. Infrared sensors are used to monitor key points in the transfer system. The sample capsule can be automatically removed from the “rabbit” in a special switch prior to spectrum collection. γ-Spectra are acquired by a HPGe detector with the Loss-Free-Counting method of Westphal. The delayed neutrons are measured by five 3He-detectors and a multichannel scaler. The transfer time is about 200 ms. The system and its mode of operation for routine work will be described.
Keywords :
Short time irradiation , Rabbit system , Delayed neutron counting , NAA
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Serial Year :
2004
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Record number :
2198199
Link To Document :
بازگشت