Characterization of the Hamamatsu R11410-10 3-in. photomultiplier tube for liquid xenon dark matter direct detection experiments
Author/Authors :
Lung، نويسنده , , K. and Arisaka، نويسنده , , K. and Bargetzi، نويسنده , , A. and Beltrame، نويسنده , , Joseph P. and Cahill، نويسنده , , A. and Genma، نويسنده , , T. and Ghag، نويسنده , , C. and Gordon، نويسنده , , D. and Sainz، نويسنده , , J. and Teymourian، نويسنده , , A. and Yoshizawa، نويسنده , , Y.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2012
Pages :
8
From page :
32
To page :
39
Abstract :
To satisfy the requirements of the next generation of dark matter detectors based on the dual phase TPC, Hamamatsu, in close collaboration with UCLA, has developed the R11410-10 photomultiplier tube. In this work, we present the detailed tests performed on this device. High QE ( ∼ 30 % ) accompanied by a low dark count rate (50 Hz at 0.3 PE) and high gain (1×107) with good single PE resolution have been observed. A comprehensive screening measurement campaign is ongoing while the manufacturer quotes a radioactivity of 20 mBq/PMT. These characteristics show the R11410-10 to be particularly suitable for the forthcoming zero background liquid xenon detectors.
Keywords :
Photomultiplier tubes , Dark matter , Astroparticle physics