Abstract
All available measured data for the response characteristics of the Leake counter have been gathered together. These data, augmented by previously unpublished work, have been compared to Monte Carlo simulations of the instrumentʹs response characteristics in the energy range from thermal to 20 MeV. A response function has been derived, which is recommended as the best currently available for the instrument. Folding this function with workplace energy distributions has enabled an assessment of the impact of this new response function to be made. Similar work, which will be published separately, has been carried out for the NM2 and the Studsvik 2202D neutron area survey instruments.
Keywords
Neutron area survey meters , neutron dosimetry , Fluence response functions , MCNP , Monte Carlo modelling , Ambient dose equivalent