Title of article :
Evaluation and Monte Carlo modelling of the response function of the Leake neutron area survey instrument
Author/Authors :
Tagziria، نويسنده , , H. G. Tanner، نويسنده , , R.J. and Bartlett، نويسنده , , D.T. and Thomas، نويسنده , , D.J.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2004
Pages :
11
From page :
596
To page :
606
Abstract :
All available measured data for the response characteristics of the Leake counter have been gathered together. These data, augmented by previously unpublished work, have been compared to Monte Carlo simulations of the instrumentʹs response characteristics in the energy range from thermal to 20 MeV. A response function has been derived, which is recommended as the best currently available for the instrument. Folding this function with workplace energy distributions has enabled an assessment of the impact of this new response function to be made. Similar work, which will be published separately, has been carried out for the NM2 and the Studsvik 2202D neutron area survey instruments.
Keywords :
Neutron area survey meters , neutron dosimetry , Fluence response functions , MCNP , Monte Carlo modelling , Ambient dose equivalent