Title of article
Novel electronics for MCP imaging detectors
Author/Authors
Lees، نويسنده , , J.E. and Bassford، نويسنده , , D. and Whitford، نويسنده , , C. and Blake، نويسنده , , O.، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2005
Pages
5
From page
621
To page
625
Abstract
We describe novel electronics for processing signals from a microchannel plate detector using a resistive anode readout. High sampling-rate (MHz) Analogue-to-Digital Converters (ADCs) coupled to a Digital Signal Processor (DSP) combine to offer high spatial resolution ( 70 μ m ) , low noise ( < 0.1 count cm - 1 s - 1 ) imaging over a large ( 93 mm × 93 mm ) imaging area.
new low-cost electronics form an integral part of a MCP-based system for digital autoradiography designed to image radiolabelled thin tissue slices and 1- and 2-D electrophoresis gels.
Keywords
MCPS , autoradiography , Resistive anodes , X-rays
Journal title
Nuclear Instruments and Methods in Physics Research Section A
Serial Year
2005
Journal title
Nuclear Instruments and Methods in Physics Research Section A
Record number