Title of article
Determination of the isotopic ratio 234U/238U in liquid scintillation and environmental applications
Author/Authors
Bianchi، نويسنده , , D. and Cortese، نويسنده , , P. and Dellacasa، نويسنده , , G. and Merlano، نويسنده , , I.، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2005
Pages
7
From page
543
To page
549
Abstract
We describe a procedure to measure the isotopic ratio 234U/238U in liquid samples. This method combines an extractive technique, liquid scintillation counting with α / β discrimination and a fit procedure of α ADC spectra. Its main advantages are: high detection efficiencies for α emissions, energy sensitivity, rapid and easy preparation of the samples. This makes the technique suitable for environmental survey. We tested the method on a wide set of NBS standards, considering 235U enrichments up to 5% (the maximum value allowed for civil uses). Results on samples collected on field were validated by a comparison with ICP mass spectrometry.
Keywords
Liquid scintillation , Uranium isotopic composition , Environmental contamination , Multi-parameter fit
Journal title
Nuclear Instruments and Methods in Physics Research Section A
Serial Year
2005
Journal title
Nuclear Instruments and Methods in Physics Research Section A
Record number
2199446
Link To Document