Title of article :
LaCl3(Ce) scintillation detector applications for environmental gamma-ray measurements of low to high dose rates
Author/Authors :
Tsutsumi، نويسنده , , Masahiro and Tanimura، نويسنده , , Yoshihiko، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2006
Pages :
7
From page :
554
To page :
560
Abstract :
A new cerium-doped LaCl3(Ce) scintillator is evaluated with respect to the application in environmental gamma-ray dosimetry and spectrometry. The scintillator is very attractive for gamma-ray spectrometry in the case of high count rate, because it has excellent energy resolution and fast decay time. The performance characteristics of a scintillator with a 25.4 mm×25.4 mm LaCl3(Ce) crystal are studied and compared to those of a NaI(Tl) scintillator with the same size crystal. Acquired pulse height spectra are converted to dose rates by using the G ( E ) function method. Though the LaCl3(Ce) crystal itself produces a rather high background in the crystal itself, the scintillator provides good energy information and dose-rate readings from low to high-level (several mGy/h) by subtracting the self-background. The properties of LaCl3(Ce) scintillator suggest that the scintillator could be a promising candidate for monitoring at high dose levels as in emergencies, as well as at ordinary levels of background radiation.
Keywords :
LaCl3(Ce) , gamma-rays , Spectrometry , dosimetry , G(E) function , NaI(Tl)
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Serial Year :
2006
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Record number :
2199701