Title of article :
Liquid-filled ionization chamber temperature dependence
Author/Authors :
Franco، نويسنده , , L. and Gَmez، نويسنده , , F. and Iglesias، نويسنده , , A. and Pardo، نويسنده , , J. and Pazos، نويسنده , , A. and Pena، نويسنده , , J. and Zapata، نويسنده , , M.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2006
Abstract :
Temperature and pressure corrections of the read-out signal of ionization chambers have a crucial importance in order to perform high-precision absolute dose measurements. In the present work the temperature and pressure dependences of a sealed liquid isooctane filled ionization chamber (previously developed by the authors) for radiotherapy applications have been studied.
e analyzed the thermal response of the liquid ionization chamber in a ∼ 20 ∘ C interval around room temperature. The temperature dependence of the signal can be considered linear, with a slope that depends on the chamber collection electric field. For example, a relative signal slope of 0.27 × 10 - 2 K - 1 for an operation electric field of 1.67 × 10 6 V m - 1 has been measured in our detector. On the other hand, ambient pressure dependence has been found negligible, as expected for liquid-filled chambers.
ermal dependence of the liquid ionization chamber signal can be parametrized within the Onsager theory on initial recombination. Considering that changes with temperature of the detector response are due to variations in the free ion yield, a parametrization of this dependence has been obtained. There is a good agreement between the experimental data and the theoretical model from the Onsager framework.
Keywords :
Free ion yield , temperature dependence , Onsager theory , isooctane , Liquid-filled ionization chamber , Absolute dosimetry
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Journal title :
Nuclear Instruments and Methods in Physics Research Section A